Calibration Kits
Calibration kits are available for the device. The recommended calibration kits are shown in the following table.
Table 4
Gas Type | CAL Cylinder | Cylinder P/N | Span Value | CAL KIT # |
---|---|---|---|---|
Methane | 2.5% Methane | 10177546 | 57% LEL (LEL=4.4% vol) | 40 |
Propane | 0.6% Propane | 10028034 | 35% LEL (LEL=1.7% vol) | 40 |
Zero Gas | 100% Nitrogen | 10188082 | 0% LEL | 40 |
See https://msasafety.info/PrimaXIRPlus for additional PrimaX IR+ calibration gases.